Lucideon Offers Accelerated Aging of UHMWPE Orthopaedic Implant Components

Lucideon has added accelerated aging of Ultra High Molecular Weight Polyethylene (UHMWPE) polymer orthopedic implants to its range of wear and fatigue testing services.

The service, which is carried out in accordance with ASTM F2003, is performed to support FDA and regulatory submissions, as well as R&D programs on new polymer materials. Both conventional polymers and advanced UHMWPE components, such as Vitamin E stabilized components, can be treated.

“The test we perform simulates the effects of real life oxidation on polymer components via raised temperature and oxygen pressures,” said Gemma Budd, Healthcare Business Manager at Lucideon. “As the oxidation may affect the mechanical performance of implants and devices, we may advise clients to have their implants aged before additional mechanical wear testing is carried out, depending on the materials being used. This simulates a worst-case scenario for evaluation.”

Typical applications for the new test include aging of acetabular liners in advance of total hip joint wear testing (ISO 14242) and impingement testing (ASTM F2582), and aging of tibial inserts in advance of total knee joint wear testing (ISO 14243).

Source: Lucideon